Mammographic Unit Assembly |
all controls functioning normally |
Grid | no lines |
Radiation Output at 30 kVp, 60 cm | > 0.10 mGy/mAs |
kVp Accuracy | ± 1.0 kVp |
kVp Reproducibility | ± 0.5 kVp |
mAs Linearity | ± 10% |
Half Value Layer (HVL) Measurement at 30 kVp | 0.30 mm < HVL < 0.4 mm A1 |
Collimation/Beam Alignment | < 2 % of SID but ≤ 5 mm at chest wall |
AEC kVp Compensation, Minimum Response Time & Backup Timer | ± 10% O.D. over 25-30 kVp |
AEC Object Thickness Compensation | ± 10% O.D. over 2-6 cm |
Image Quality with RMI 156 Phantom (minimum) |
0.75 mm fibres
0.32 mm specks
0.75 mm mass
|
Irradiation Coefficient of Variation | 5 % |
Lucite Block Density | 1.2 ± 0.1 O.D. |
Breast Surface Dose | < 9 mSv |
Average Glandular Dose | < 3 mSv per view |