APPENDIX E

QUALITY CONTROL TESTING AND MAMMOGRAPHIC EOUIPMENT STANDARDS

A documented quality control program with procedure manuals and logs shall be maintained and be available to all staff. The quality control program shall include the following tests as a minimum16-19,21-25.

Radiation Technologist's Checks: Minimum Frequency
Darkroom CleanlinessDaily
Processor SensitometryDaily
Screen CleanlinessWeekly
Viewboxes and Viewing ConditionsWeekly
Image Quality with RMI 156 PhantomWeekly
Visual Check ListMonthly
Screen UniformityQuarterly
Darkroom Fog Tests (non-daylight systems)Semi-annually
Film/Screen Contact EvaluationSemi-annually
Compression Assembly EvaluationSemi-annually
Repeat Analysis AuditAnnually

Quality Assurance Technologist's/Clinical Engineer's Checks:
The following checks are to be completed semi-annually and after any servicing, with biennial audits conducted under the provincial radiation protection program. The performance criterion for each parameter is noted.

Parameter Standard/Performance Criteria
Mammographic Unit Assembly all controls functioning normally
Gridno lines
Radiation Output at 30 kVp, 60 cm> 0.10 mGy/mAs
kVp Accuracy± 1.0 kVp
kVp Reproducibility± 0.5 kVp
mAs Linearity± 10%
Half Value Layer (HVL) Measurement at 30 kVp0.30 mm < HVL < 0.4 mm A1
Collimation/Beam Alignment< 2 % of SID but ≤ 5 mm at chest wall
AEC kVp Compensation, Minimum Response Time & Backup Timer± 10% O.D. over 25-30 kVp
AEC Object Thickness Compensation± 10% O.D. over 2-6 cm
Image Quality with RMI 156 Phantom (minimum) 0.75 mm fibres
0.32 mm specks
0.75 mm mass
Irradiation Coefficient of Variation5 %
Lucite Block Density1.2 ± 0.1 O.D.
Breast Surface Dose< 9 mSv
Average Glandular Dose< 3 mSv per view